● Evaporator Refrigerant Leakage Test
● Condenser, compressor, A/C system refrigerant
    leakage test 
● Leakage test via chamber temperature 
 
 
 
 
- Feature -
 
Test Sequence Setup 
Graphical display
PID display
Calibration
Real time graph
Maintenance
history management
Past test result display
PLC I/O monitoring &
control
1412 photoacoustic field
gas-monitor
R-134a(0.005~50PPM), Co2(1.5~150,000PPM)
Temperature setpoint (40℃, 50℃ 2Point)
   
   
   
   
- Specifications -
 
MODEL
SEOUL-SHED-3000Series
Dimension
Chamber : W2.6 x H2.8 x D6.0m (Machinery room include)
Door : W2.6 x H2.8m (Overlap type Air Packing System)
Material
Interior : SUS 304 2B Steel Plate 2.0t
Exterior : Galvanized Color Steel Plate 0.6t
Floor : SUS 304 2B Steel Plate 3.0t
Insulation
Poly Urethane Panel 50t
Viewing
Size : W0.4 x H0.7m Pair Glass 36t
Air Circulation
Motor : Mixing Motor 150W * 2ea
Blower : Sirroco Fan 3" * 5" * 2ea (SUS304)
Temp. Range
10 ~ 50'C (Accuracy ±0.5'C)
Heating Load
10,000kcal/h
Cooling Load
3,000kcal/h
Cooling System
Air Cold Condensing Chiller Unit (R-22)
Sensor
RTD PT-100Ohm
Humidity (%RH)
Brometric Pressure
Control Interface
Direct Reading / LabVIEW SoftWare with PC Interface
Power Source
AC 380V 3P 60Hz
Main Composition
SHED Chamber
AHU (Air Handring Unit)
Air Circulation System
Power Supply System
Control System & DAS
Electric Cabinet & Control Panel